The solvent keeps flowing into the titration vessel after the addition, without the add key being pressed. Set the solvent bottle up in such a way that the fluid level in the bottle is lower than in the titration vessel.
Solvent is added simultaneously during aspriation. The drying tube in the titration vessel is blocked:
- Increase hole on the drying tube cover to at least 2 mm
- Refill the drying tube, do not pack to tightly, possibly removing some of the molecular sieve.
The pump conveys no fluid. Press the addition / aspiration key more strongly.
- In most cases the bottle attachments are not screwed tightly enough to the reagent and waste bottles
- Check all connections for leak-tightness
- Check wether the 803 Ti Stand has been set up correctly
|
|
|
|